Temperature

Temperature
Topic LCI-SPG-0010, Last Updated 11/08/2024

The ambient temperature in the laboratory must be from 4 to 40 °C (39 to 104 °F). Short-term thermal variations should be no more than 2 °C (3.6 °F) per hour.

Note: Failure to operate in this range will compromise system performance and can result in instrument failure.

The optimum temperature range is 19 to 22 °C (66 to 72 °F).


Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Back To Top Back To Top