Solvent stabilizers

Solvent stabilizers
Topic LCI-USG-0121, Last Updated 11/08/2024

Solvent stabilizers are added to slow or stop solvent degradation.

Warning: Certain solvents degrade, or become unstable, over time. Highly unstable solvents present a potential explosion hazard.
Note: Do not leave solvents containing stabilizers, such as THF with butylated hydroxytoluene (BHT), to dry in the system’s flow path. A dry flow path, including the detector flow cell, becomes contaminated with residual stabilizer, and a substantial cleaning effort is needed to restore the flow path to its initial condition.

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Back To Top Back To Top