Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs! Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs!
Mobile phase solvents and the seal wash, needle wash, and purge solvents are set up before running equilibration.
Warning: To avoid personal contamination with biologically hazardous or toxic compounds, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid eye injury, use eye protection when performing this procedure.
Required tools and materials
Chemical-resistant, powder-free gloves
Protective eyewear
Clean solvent bottles
To set up solvents:
Prepare solvents in the clean bottles as required for the method.
Note: There can be up to four mobile phase solvent lines, identified by the system as A, B, C, and D. Pay particular attention to how the solvents are lettered in the project because their identifiers determine which solvent line to connect to each bottle.
The seal wash, needle wash, and purge solvents have dedicated lines, but it is possible for a method to specify filling a bottle with the same preparation as a mobile phase bottle.
If necessary, obtain the solvent specifications for the method from the SOP or the Empower project.
Rotate each of the solvent bottles:
Remove the cap and solvent line tube from the bottle that is being replaced.
Remove that bottle from the tray.
Pass the tube whose identification tag corresponds to the solvent through the opening in the replacement cap and install the cap on the bottle.
Place that bottle on the tray as shown in the following figure.
Note: If it is necessary to prime solvent lines outside of the equilibration workflow, return to the
Home screen and select the appropriate workflow on the
Setup > Solvents screen.
To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.