Consult the "Safety advisories" appendix in this publication for a comprehensive list of warning advisories and notices.
Need Help? Connect with a Waters expert.
To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.