Servicing the pump compartment air filter

Servicing the pump compartment air filter
Topic LCI-USG-0167, Last Updated 11/08/2024

You can clean or replace the pump compartment air filter.

Required tools and materials

  • Mild detergent and water
  • Air filter (if replacing)

To service the air filter:

  1. Open the pump compartment door.
  2. Compress the air filter upward above the frame relief area. Then pinch the filter and pry it away from the frame.
    Figure 1: Pump compartment air filter
    Air filter
    Air filter frame
    Frame relief area
  3. Do one of the following:
    • Clean the air filter using a mild detergent and water, and then dry the filter.
    • Discard the old air filter.
  4. Compress the air filter slightly and reattach inside the frame.
  5. Close the pump compartment door.