General guidelines

General guidelines
Topic LCI-USG-0044, Last Updated 11/08/2024

When performing an HPLC analysis, follow these general recommendations.

  • Use high-quality (HPLC- or MS-grade) solvents, buffers, and additives.
  • Use high-quality (HPLC- or MS-grade) water.
  • Always use solvent filters on tubing in solvent bottles.
  • Filter buffers with a 0.45-µm filter membrane.
  • Keep concentrated stock solutions to use when preparing working solutions, and, to maximize shelf life, refrigerate them when not using them.
  • Do not add fresh buffer to old (a practice known as "topping off"). Doing so can promote microbial growth.
  • Keep all solvent lines primed.
  • Flush buffers from the system when they are not in use, taking care to avoid using solvents that can precipitate or otherwise react.
  • Use 10%-20% organic solvent in water as a storage solvent if you expect the system to remain idle longer than 24 hours.
  • Keep the seal wash line primed.
  • Monitor the level in the waste vessel to ensure that it can accommodate all the expected waste.

For information about preventing and eliminating contamination, see Controlling Contamination in LC/MS Systems (715001307) on the Waters website (www.waters.com).

Contents

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Back To Top Back To Top