TUV detector optics

TUV detector optics
Topic LCI-USG-0112, Last Updated 11/08/2024

The detector optics are based on a Fastie-Ebert monochromator.

The following figure shows the components that make up the detector optics assembly.

Note: In addition to what is shown in the figure, detector optics also include a shutter, wavelength calibration filter, and second-order filter.
Figure 1: Detector optics assembly
Mirror
High-brightness deuterium (D2) lamp
Window
Filter wheel
Slit
Mirror
Grating
Reference photodiode
Beam splitter
TaperSlit flow cell
Sample photodiode

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Back To Top Back To Top