Injection mechanics

Injection mechanics
Topic LCI-USG-0068, Last Updated 11/08/2024

The sample needle goes through a specific sequence for each injection.

During an injection, the following sequence takes place:

  1. The needle moves to the sample vial and aspirates sample from it.
  2. The needle carriage inserts the needle into the injection/wash port.
    Note: The entire injection port assembly rests on a spring and is guided inside a metal housing.
  3. As the needle is inserted into the injection/wash port, it presses against the seat and forms a high-pressure seal.
  4. The injection valve turns to start the injection.
  5. The wash pump washes the outside of the needle during the injection.
Figure 1: Needle seat assembly, cross view
Angled wash manifold
Sample needle support sleeve
Fitting
Seat
Seat port assembly
Seal cup locking nut
Compression spring
Needle wash housing

The following figure shows standard injection mode cycle time.

Figure 2: Standard injection mode cycle time definition

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Back To Top Back To Top