PDA optics assembly light path

PDA optics assembly light path
Topic LCI-USG-0219, Last Updated 11/07/2024

The detector employs an extremely efficient design for exceptionally high light throughput.

The detector operates as follows:

  1. The light emitted by the deuterium lamp is refocused by an elliptical mirror into the flow cell.
  2. The light beam then expands to fill the grating, which separates the beam into its component wavelengths and then refocuses those onto the PDA sensor.

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Back To Top Back To Top