Wash solvent recommendation

Wash solvent recommendation
Topic LCI-USG-0128, Last Updated 11/08/2024

To maintain system condition and overall performance, follow this wash solvent recommendation.

The use of nonvolatile (solid salt) buffer solutions in wash solvents is permitted in limited concentration only. Generally, however, it is not recommended.

Solid, salt-based buffer solutions can dry and leave a salt residue that can scratch sealing surfaces, clog tubing, and damage the wash pump. Some applications control pH to promote sample solubility, which manifests in better peak shape and narrower peaks upon elution. Better sample solubility in the mobile phase also results in the sample components being less likely to adhere onto surfaces such as the stationary phase or the needle. If pH control is important to separation performance, consider it when determining the composition of wash solvent. For example, if acid (low pH) is required to keep the sample in solution during the separation, it is likely to be a required component of the wash solvent, necessary to dissolve the sample adhering to the needle's surface and to rinse the wash station.

Caution: When operating the Alliance iS Bio HPLC System above pH 10, ensure that you are using the optional high pH kit. For further assistance, contact Waters.

Need Help? Connect with a Waters expert. 


To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.

Back To Top Back To Top