To prevent local power instability from impacting system reliability and performance, Waters recommends using an uninterruptible power supply (UPS). In support of this recommendation, Waters supplies UPS systems configured and evaluated specifically for use with Waters MS systems. These UPS units can step up single-phase line voltage to 230 V ac, providing power conditioning and protection for the MS and LC system components.
For North America, the UPS system requires one L6-30, 30-amp wall socket. In other areas, the UPS system typically connects to laboratory main power using the standard power cord and wall socket required for your instrument system.
需要協助嗎?聯絡Waters專家。
為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。