Power outlet requirements

Power outlet requirements
Topic LCI-SPG-0002, 最近更新 09/27/2023

The system is shipped with the power cords requested at the point of order. The customer is responsible for ensuring that their power outlets are compatible with the power cords shipped with the product. The power cord sets must comply with local regulations. To ensure electrical safety, use only the power cord set included with your Waters product. Do not use the Waters-provided power cord for any other products. The power cord set is defined as the cable with plugs at both ends.

Note: If you want to install ancillary equipment (for example, compressors), you may need additional power outlets, possibly requiring 3-phase supplies. You must confirm such supplemental needs with the local Waters agent prior to the start of the installation.

To help determine the receptacles required for the components in your system, refer to the following table.

Table 1. Regional power cords supplied by Waters
Region

Power connection (IEC 60320 C19 [16A rating])

US/Canada

NEMA 5-15P

Australia

15A

Brazil

16A

China

16A

Denmark

DK 2-1a; 13A

Europe

CEE 7/VII "Schuko"; 16A

India

16A

Japan

PSE approved JIS C 8303 15A 3 prong type B plug

Korea

CEE 7/VII "Schuko"; 16A

Switzerland

Type 23; 16A

Taiwan

5-15; 15A

UK

13A