Follow all local electrical safety requirements in preparing your laboratory.
Notes:
- Installations in Australia and New Zealand must comply with AS3000: Electrical Installations for Australia and New Zealand.
- Each branch circuit supplying power to the system and accessories must provide a protective earth ground and short circuit protection using a circuit breaker or a fuse rated for the load in accordance with the country's national electric code and/or local regulations. Multiple branch circuits are required.
Warning: To avoid electric shock, all system components require a dedicated, earthed (grounded) power source. The receptacles from this power source must be accessible to the system components, and they must share a common ground.
Notes: Waters recommends providing additional protection by means of the following:
- Residual current devices (RCDs) for UK and Europe
- Ground fault circuit interrupters (GFCIs) for the rest of the world