Mercury calibration uses a mercury-argon spectral calibration lamp to calibrate or recalibrate the instrument. Mercury calibration is typically performed at the factory but can be performed by qualified Waters field service engineers if necessary. Note that mercury recalibration should only be performed if erbium calibration fails or the prior mercury calibration parameters were wiped or corrupted.
A Waters field service engineer can perform a mercury calibration for the PDA detector from the system touchscreen.
Need Help? Connect with a Waters expert.
To avoid risk of personal injury and to prevent damage to laboratory equipment, always operate your Waters products in accordance with the applicable operational and safety information, your organization’s standard operating procedures, and local regulations.