Solvent stabilizers

Solvent stabilizers
Topic LCI-USG-0121, 最近更新 11/28/2023

Solvent stabilizers are added to slow or stop solvent degradation.

Warning: Certain solvents degrade, or become unstable, over time. Highly unstable solvents present a potential explosion hazard.
Note: Do not leave solvents containing stabilizers, such as THF with butylated hydroxytoluene (BHT), to dry in the system’s flow path. A dry flow path, including the detector flow cell, becomes contaminated with residual stabilizer, and a substantial cleaning effort is needed to restore the flow path to its initial condition.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首