Pump features

Pump features
Topic LCI-USG-0051, 最近更新 11/28/2023

The system uses a low-pressure mixing pump.

The Alliance iS Quaternary Solvent Manager (QSM) can blend up to four degassed solvents (A, B, C, and D). A gradient proportioning valve (GPV) is used to dynamically blend solvents in any specified combination, producing repeatable and reproducible gradient segments regardless of solvent compressibility and system back pressure. Integrated degasser chambers (one per solvent line) automatically remove dissolved gases from up to four elution solvents.

The pump features:

  • Automated, continuous compressibility compensation for accurate and precise delivery at pressures up to 10,000 psi
  • Leak sensors to identify and manage solvent leaks during unattended operation
  • Programmable flow rates that range from 0.001 to 10.000 mL/min, in 0.001-mL increments

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首