For the system's sample manager, consult this table of injection volumes and their corresponding expected precision (area reproducibility) measurements.
Criteria | Value | Minimum volume | Maximum volume |
---|---|---|---|
% Relative standard deviation (RSD) | ≤1.00 | ≥0.5 µL | <1.0 µL |
≤0.50 | ≥1.0 µL | <5.0 µL | |
≤0.25 | ≥5.0 µL | <50.0 µL |
需要協助嗎?聯絡Waters專家。
為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。