Installing the column

Installing the column
Topic LCI-USG-0007, 最近更新 11/28/2023

Install the column in the CHC prior to running samples.

Tool-free fittings and column clips are designed to be intuitive when installing a column in the Alliance iS Column Heater Cooler (CHC).

Note: Power-on the system before installing the column to ensure proper RF column identification when the CHC door is closed. See Powering-on the system.
  1. Open the column compartment door.
  2. Move the lower column clip as necessary to match the size of the column.
  3. Remove the plugs from the inlet and outlet ends of the column.
  4. Orient the column so that the outlet faces up (see the arrow on the column) and the inlet faces down.
  5. Screw the tool-free fittings on the column compartment tubing finger-tight to the column inlet and outlet.
  6. Insert the column into the upper and lower clips so that the clip grasps the exposed threads on the tool-free fitting.
    Figure 1: Column seated in column clip
  7. Close the column compartment door.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首