Starting the hardware and software

Starting the hardware and software
Topic LCI-USG-0171, 最近更新 11/28/2023

When the Alliance iS HPLC System hardware is running, start the Empower software.

To start the hardware and software:

  1. Unlock or power-on the Alliance iS HPLC System. (If the system is powered-down, see Powering-on the system.)
  2. From the touchscreen, tap the Commands button in the left-hand pane.
    The Commands view appears.
  3. Ensure that Detector Lamp status is Lamp is on or Lamp is warming.
    If the status is Lamp is off, tap the bulb button and hold it as the timer counts down. Then wait for the status to progress from Lamp is warming to Lamp is on.
    Note: Allow 30-60 minutes for the detector lamp to warm up. For this reason, using the Detector Lamp command to turn on the lamp is preferable to turning it on during the setup process.
  4. Start Empower at the workstation.
  5. Open Run Samples.
    The Empower control panel starts automatically in the status pane.
  6. From the control panel, start the Alliance iS HPLC System console by clicking the right-arrow in the upper-right corner.

    From the console, you can access configuration, diagnostics, and detailed statuses for all parts of the system.

  7. If necessary, open the Empower project needed for the analysis.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首