Installing an extension loop in a single-valve system

Installing an extension loop in a single-valve system
Topic LCI-USG-0129, 最近更新 11/28/2023

In a single-valve system, the extension loop is installed between the pressure transducer and the sample needle.

  1. If system flow is running, stop the flow. On the touchscreen, tap COMMANDS, and then tap the pause button next to Flow is On.
  2. Open the sample manager fluidics door.
    Figure 1: Single-valve system
    Extension loop with holder
    Sample needle tool-free fitting
    Pressure transducer
    TFF adapter fitting
  3. Disconnect the tool-free fitting (TFF) that is attached to the extension loop and the sample needle. There is no need to remove the TFF adapter fitting from the pressure transducer.
    Figure 2: Extension loop latch
    Extension loop latch
  4. Push the extension loop latch backward to release the extension loop holder and remove it from the sample manager.
  5. Install a different size extension loop between the TFF adapter fitting and the sample needle. There is no need to remove the TFF adapter fitting from the pressure transducer.
  6. Push the extension loop latch forward to secure the extension loop holder in the sample manager.
  7. See Modifying needle and extension loop configuration parameters, for reporting the correct extension loop volume configuration on the touchscreen.
  8. On the touchscreen, tap Setup > Startup and follow the remaining on-screen prompts to prime and prepare the system for use.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首