Recovering maximum sample from vials

Recovering maximum sample from vials
Topic LCI-USG-0086, 最近更新 11/28/2023

Improper system settings can negatively affect how much sample is drawn from the vial.

The system has an optional vial bottom sensing feature. When it is enabled, the system will probe the bottom of the vial and then back off slightly so that the needle is very close to the bottom.

The default ANSI plate (48 vials) definition for the 2-mL Maximum Recovery Vials can leave some sample in the vial. If you must recover the maximum amount of sample possible, turn on the vial bottom sensing function.

See also: For information about enabling the vial bottom sensing feature, see Setting up a method.
See also: Waters Sample Vials and Accessories brochure on www.waters.com.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首