Transferring methods

Transferring methods
Topic LCI-USG-0123, 最近更新 11/28/2023

The Alliance iS HPLC System can produce results comparable to those achieved on many other HPLC systems. However, when moving a regulated method from one system to another, typically no modifications to the instrument method can be made. For example, the column type (diameter) specified for a transferred method cannot be changed. Even with the same column type, results may not always be identical. If there is a significant difference in retention time, the dwell volume can be compensated by adjusting the gradient start relative to the injection. Extra-column effects can also be modified.

See the Dwell Volume and Extra-Column Volume: What Are They and How Do They Impact Method Transfer white paper (720005723EN) on www.waters.com.


需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首