Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs! Waters offers a comprehensive range of analytical system solutions, software, and services for scientists. Liquid Chromatography. Mass Spectrometry. Waters is the leading provider of lab equipment, supplies and software for scientists across the world. Easily research and order everything your lab needs!
Users or Waters field service personnel can replace the pump mixer.
Warning: Observe Good Laboratory Practice (GLP) at all times, particularly when working with hazardous materials. Consult the Safety Data Sheets regarding the solvents you use. Additionally, consult the safety representative for your organization regarding its protocols for handling such materials.
Note: To avoid contaminating system components, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Required tools and materials
Chemical-resistant, powder-free gloves
1/4-inch open-end wrench
3/8-inch open-end wrench
Replacement mixer
To replace the mixer:
Flush the pump with nonhazardous solvent.
Stop the solvent flow.
Open the pump compartment door.
Remove the mixer from the mixer clip.
Using the 3/8-inch open-end wrench to hold the mixer in place, disconnect the outlet compression fitting using the 1/4-inch open-end wrench.
Mixer
Place the 3/8-inch open-end wrench here (on the flat surface)
Outlet compression fitting
Using the 3/8-inch open-end wrench to hold the mixer, disconnect the inlet compression fitting using the 1/4-inch wrench.
Inlet compression fitting
Place the 3/8-inch open-end wrench here (on the flat surface)
Mixer
Unpack the replacement mixer.
Note: Ensure that the arrow on the mixer is pointing from left to right.
Reattach the compression fittings to the mixer and tighten them finger-tight, plus as much as an additional 1/6-turn for existing fittings or 3/4-turn for new fittings.