Data acquisition checks

Data acquisition checks
Topic -0182, 最近更新 11/28/2023

Lab managers can set up pre-run and run-time data acquisition checks for minimizing common errors. The touchscreen provides the interface for setting the checks. From the Home screen, the path to the workflows is System > Administration > Acquisition Checks. The Acquisition Checks screen offers Pre-Run Checks and Run-Time Checks.

Note: This feature works with Empower 3.8.0 and later.

Data acquisition can proceed when all active pre-run checks pass. Checks can include:

  • Column must be installed: Checks whether a column is installed in the column heater, but only if using a Waters column with a readable tag.
  • Column must match method: Checks whether the installed column is appropriate for the selected method, but only if using a Waters column with a readable tag.
  • No pending preventative maintenance: Checks the date set by Preventative Maintenance > Set Preventative Maintenance.
  • System is qualified: Checks the date set by Administration > System Qualification > Set the System Qualification. This date must be supported by valid documentation and test results and is typically updated when a Waters engineer re-qualifies the system. Data acquisition cannot proceed if qualification is expired.
  • Mobile phase is not expired: Checks the date for each of the mobile phase solvents (A, B, C, D) used by the method.
  • Sample plates must be installed
  • Sample plates must match method
  • All vials present

Data acquisition halts when run-time checks detect selected issues while sample sets are running. Checks can include:

  • Mobile phase is low: When any mobile phase solvent bottle is less than 10% full.
  • Wash solvent is low: When any wash solvent bottle is less than 10% full.
  • Leak is detected: Always enabled. Checking of the pump, column, TUV detector, or Sample Manager leak sensors is controlled by System > Leak Sensors.
  • Vial is missing: Always enabled. Checking for vials in the locations specified for the sample set is automatic.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首