Measuring dwell volume

Measuring dwell volume
Topic LCI-USG-0110, 最近更新 11/28/2023

When transferring a gradient LC method, you can produce comparable retention times by measuring dwell volume on both systems. The dwell volume is the system volume between the point where the gradient is formed and the column inlet.

You can measure dwell volume using the midpoint of a 0–100% gradient. To do so, run a gradient between two identical solvents, A and B, spiking the B solvent with a marker. Perform the measurement after configuring the system without the column for the instrument method being transferred, replacing the column with a low-volume restrictor to ensure proper pump function.

See also: “Measuring system volume for methods transfer” in the Empower online Information System.


需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首