Use miscibility numbers to predict the miscibility of a liquid with a standard solvent.
To predict the miscibility of two liquids, subtract the smaller miscibility number (M-number) value from the larger M-number value:
Some solvents prove immiscible with solvents at both ends of the lipophilicity scale. These solvents receive a dual M-number:
For example, some fluorocarbons are immiscible with all the standard solvents and have M-numbers of 0 and 32. Two liquids with dual M-numbers are usually miscible with each other.
A liquid is classified in the M-number system by testing for miscibility with a sequence of standard solvents. A correction term of 15 units is then either added or subtracted from the cutoff point for miscibility.