Priming the system

Priming the system
Topic LCI-USG-0043, 最近更新 11/28/2023

After powering-on the system, you must prime it before the system is ready for use.

Requirement: You must prime the system after starting it, as well as after changing the mobile phase, after changing the sample needle, and after the system is idle for four hours or more.
Requirement: You must install a column before priming the system. See Installing the column.
Recommendation: If you are introducing new solvents, prime them at 4 mL/min for seven minutes. Alternatively, prime the solvents at 4 mL/min for three minutes. Ensure that sufficient quantities of solvent are available for priming.

There are multiple ways to prime the system from the touchscreen:

  • Tap SETUP > Startup to prime all solvents, prime the needle wash and seal wash, and specify the solvent composition, flow rate, column and sample temperatures, and needle characterization for your next system startup.
  • Tap SETUP > Solvents > Prime Mobile Phase Solvents to prime the pump.
  • Tap HOME, tap the solvent bottle icon, tap a mobile phase condition card, and then tap Prime Solvent to prime an individual mobile phase.
  • Tap SETUP > Solvents > Prime Sample Metering Pump to prime the sample metering pump.
Tip: From the touchscreen, you can select the Setup > Startup feature to prime all solvents, prime the needle wash and seal wash, and specify the solvent composition, flow rate, column and sample temperatures, and needle characterization for your next system startup. For details, see the touchscreen.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首