Replacing the column heater leak sensor

Replacing the column heater leak sensor
Topic LCI-USG-0188, 最近更新 11/28/2023

Users or Waters field service personnel can replace the column heater sensor.

Warning: Observe Good Laboratory Practice (GLP) at all times, particularly when working with hazardous materials. Consult the Safety Data Sheets regarding the solvents you use. Additionally, consult the safety representative for your organization regarding its protocols for handling such materials.
Warning: To avoid personal contamination with biologically hazardous or toxic compounds, wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Warning: To avoid eye injury, use eye protection when performing this procedure.
Requirement: Wear clean, chemical-resistant, powder-free gloves when performing this procedure.

Required tools and materials

  • Chemical-resistant, powder-free gloves
  • Protective eyewear
  • Replacement leak sensor

To replace the leak sensor:

  1. Open the column compartment door.
  2. Press down on the tab to detach the leak sensor connector from the front of the device.
    Figure 1: Leak sensor connector
    Tab
    Leak sensor connector
  3. Grasp the leak sensor by its serrations and pull upward on it to remove it from its reservoir.
    Figure 2: Leak sensor serrations
    Serrations
  4. Unpack the new leak sensor.
  5. Align the leak sensor’s T-bar with the slot in the side of the leak sensor reservoir, and then slide the leak sensor into place.
    Figure 3: Aligning T-bar with slot
    T-bar
    Slot in leak sensor reservoir
    Leak sensor port on front of device
  6. Connect the leak sensor connector to the front of the device.
    Figure 4: Attaching leak sensor connector
    Leak sensor connector
    Leak sensor port on front of device
  7. Close the column compartment door.
  8. From the touchscreen Commands view, tap Reset.
  9. From the touchscreen System view, tap Leak Sensors and then enable the QSM Leak Sensor.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首