Replacing the detector leak sensor

Replacing the detector leak sensor
Topic LCI-USG-0177, 最近更新 11/28/2023

A leak sensor in the drip tray continuously monitors the detector for leaks.

Warning: To avoid personal contamination with biologically hazardous or toxic compounds, wear clean, chemical-resistant, powder-free gloves when performing this procedure.

The detector's leak sensor stops system flow when it detects accumulated, leaked liquid in its reservoir. When the sensor detects a leak, the system's touchscreen displays an alarm message.

Required tools and materials

  • Chemical-resistant, powder-free gloves
  • Replacement leak sensor

To replace the detector's leak sensor:

  1. Open the detector's door, gently pulling its right-hand edge toward you.
  2. Remove the old leak sensor:
    1. To detach the leak sensor connector from the front of the instrument, press down on the release tab.
      Note: The following figure is for representation only. Your hardware may differ slightly from that shown.
      Figure 1: Detaching the leak sensor
      Release tab
      Leak sensor connector
    2. To remove the leak sensor from the reservoir, grasp it by the serrations around its top and pull upward on it (see Figure "Installed leak sensor, rear view").
  3. Install the new leak sensor:
    1. Holding the leak sensor by its serrations, align its T-bar with the slot in the side of the reservoir and slide it into place (see Figure "Installed leak sensor, rear view").
      Figure 2: Installed leak sensor, rear view
      Serrations
      T-bar engaged with reservoir slot
      Figure 3: Installed leak sensor, front view
      Leak sensor
    2. Plug the leak sensor connector into the front of the instrument.
  4. Close the detector's door.
  5. From the system's touchscreen Commands view, tap Reset.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首