Calibrating the needle z axis

Calibrating the needle z axis
Topic LCI-USG-0181, 最近更新 11/28/2023

Users or Waters field service personnel can calibrate the sample manager needle's z axis.

You must calibrate the needle before you use the sample manager for the first time and whenever you replace the sample needle. Failing to calibrate the needle can damage it. The calibration procedure is identical for all needles.

To calibrate the needle z axis, do the following:

  1. From the touchscreen, tap Maintain > Calibrate needle Z axis.
  2. Follow the directions on the screen.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首