Solvents to avoid

Solvents to avoid
Topic LCI-USG-0154, 最近更新 11/28/2023

Consult this list of solvents to avoid using in your system.

Avoid the following substances:

  • Solvents that contain halogens:
    • Bromine
    • Chlorine
    • Fluorine
    • Iodine
  • Warning: Peroxide contaminants in THF can spontaneously and destructively explode when you partially or completely evaporate the THF.
    Compounds that form peroxides, such as ultraviolet (UV)-grade ethers, non-stabilized THF, dioxane, and diisopropylether (If you must use compounds that form peroxides, be sure to filter them through dry aluminum oxide, to adsorb formed peroxides. Do not allow them to remain on a system for more than 24 hours.)
  • Solutions that contain strong concentrations (greater than 0.1% wt) of complexing agents like EDTA
  • Note: The standard pH operating range for your system is 1.0 to 13.0. Operating the system below pH 1.0 or above pH 13.0 for more than brief periods can result in increased wear on system components not included in preventive maintenance kits and the need for more frequent routine preventive maintenance.
    Strong acids and strong bases
    • For strong acids, unless using them as cleaning agents, use them only in weak concentration. Avoid using acids as mobile phases when their pH is less than 1.0.
    • For strong bases, use them only in weak concentration. Avoid using bases as mobile phases when their pH is greater than 13.0.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首