Moving the pump plunger backward

Moving the pump plunger backward
Topic LCI-USG-0195, 最近更新 11/28/2023

Use the touchscreen UI to move the pump plunger backward before replacing pump components.

To move the pump plunger backward:

  1. Flush the pump with nonhazardous solvent.
  2. From the touchscreen, select the solvent manager, and then tap Maintain > Replace Components.
  3. From the Replace Components page, scroll down and then tap Pump Maintenance.
  4. Tap Pump Head Maintenance.
  5. Select the plunger head you want to move backward.
  6. Tap Move Backward, and then wait for the plunger to stop.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首