Short or long 1/4-28 fitting with flangeless ferrule and stainless steel lock ring

Short or long 1/4-28 fitting with flangeless ferrule and stainless steel lock ring
Topic LCI-USG-0067, 最近更新 11/28/2023

Tighten the fitting finger-tight.

Important: Short fittings are intended for use with 1/16-inch OD tubing. Long fittings are intended for use with 1/8-inch OD tubing.

Figure 1: Short 1/4-28 fitting with flangeless ferrule and SST lock ring, first use or reinstallation
Compression screw
Lock ring
Ferrule
End of lock ring with larger ID
Figure 2: Long 1/4-28 fitting with flangeless ferrule and SST lock ring, first use or reinstallation
Compression screw
Lock ring
Ferrule
End of lock ring with larger ID

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首