Dual-Wavelength mode

Dual-Wavelength mode
Topic LCI-USG-0117, 最近更新 11/28/2023

In Dual-Wavelength mode, the detector can monitor two wavelengths, one on channel A and one on channel B.

Because the sampling frequency is reduced to 1 or 2 Hz, use of Dual-Wavelength mode is limited to more standard chromatography where peak widths span at least 20 seconds to enable full characterization of a peak.

In this mode, the following conditions apply:

  • If both selected wavelengths are greater than 370 nm, the detector applies the second-order filter to block unwanted UV light.
  • If both selected wavelengths are less than or equal to 370 nm, the detector removes the second-order filter.
  • If the selected wavelengths bracket the 370-nm threshold, the detector does not apply the second-order filter and issues an alarm message that any data collected for the wavelength above 370 nm can contain inaccuracies because of possible UV light interference (UV effects).

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首