Wash solvent recommendation

Wash solvent recommendation
Topic LCI-USG-0128, 最近更新 11/28/2023

To maintain system condition and overall performance, follow this wash solvent recommendation.

The use of nonvolatile (solid salt) buffer solutions in wash solvents is permitted in limited concentration only. Generally, however, it is not recommended.

Solid, salt-based buffer solutions can dry and leave a salt residue that can scratch sealing surfaces, clog tubing, and damage the wash pump. Some applications control pH to promote sample solubility, which manifests in better peak shape and narrower peaks upon elution. Better sample solubility in the mobile phase also results in the sample components being less likely to adhere onto surfaces such as the stationary phase or the needle. If pH control is important to separation performance, consider it when determining the composition of wash solvent. For example, if acid (low pH) is required to keep the sample in solution during the separation, it is likely to be a required component of the wash solvent, necessary to dissolve the sample adhering to the needle's surface and to rinse the wash station.


需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首