Column heater/cooler operation

Column heater/cooler operation
Topic LCI-USG-0139, 最近更新 11/28/2023

This module is a combination conductive heater and cooler.

When the column compartment temperature is set, either directly from the console or within a method, a command is sent to the CHC to switch the compartment heater/cooler engine on or off. On the basis of feedback from the compartment thermistor, the thermoelectric device continues to heat, or cool, until the compartment achieves the specified temperature set point.

Recommendation: When sample and column temperature are important to an application, in addition to specifying explicit temperature set points in the method, specify appropriate temperature limits. Together, these settings ensure that system operation occurs only within the defined limits and that any occurrence of an unacceptable deviation from the set points is flagged by an error message acknowledging the variance.

需要協助嗎?聯絡Waters專家。 


為避免引起人身傷害的風險和防止損壞實驗室設備,操作Waters產品時,請務必一律按照適用的操作和安全資訊,也務必遵守您所在組織的標準作業程序和當地法規。

回到頁首 回到頁首